Interpretation and evidence-review guidance · Reviewed 26 November 2026

When can I trust an antibody?

Trust is specific to the assay, specimen and clinical question. A technically convincing stain still needs evidence that it distinguishes the tumors in your differential.

Before acting on a result: confirm that the assay is validated for this material, controls are appropriate, the pattern is interpretable, and the diagnostic claim is supported in the relevant differential. Resolve a discordant result before using it to assign lineage, an alteration or treatment eligibility.

1 · Intended use

Define the exact claim

Specify the tumor or differential, specimen type, clone/platform, localization and cutoff. Distinguish a lineage clue, a molecular surrogate, a prognostic association and a treatment biomarker. Evidence for one use does not automatically validate the others.

2 · Assay validation

Check the laboratory’s validation

Confirm validation or verification for the antibody, platform, protocol and scoring system. Check whether the material matches the validation: whole section, biopsy or cytology; fixation and decalcification matter. Different clones and preparations require evidence of applicable performance.

CAP analytic-validation guideline, 2024

3 · Controls and pattern

Decide whether this stain is interpretable

Check expected internal and external controls, cellular localization, background and artifact. For a loss marker, adequate retained internal staining is essential. Failed controls make a result uninterpretable. Record focal, heterogeneous or equivocal staining and resolve it using the validated procedure.

4 · Diagnostic specificity

Look at the competing diagnoses

A series of positive target tumors establishes neither specificity nor a primary site. Examine relevant mimics, poorly differentiated variants and metastatic specimens. A high expression rate in known primaries may not describe performance in CUP.

5 · Reproducibility

Check how consistently it works

Look for transparent thresholds, blinded reading, observer agreement and reproducibility across runs or laboratories. Separate independent replication from overlapping cohorts and internal train/test splits. A large selected cohort does not remove bias.

6 · Confirmation

Resolve results that change the diagnosis

Integrate the result with morphology, other markers and clinical distribution. Use orthogonal or molecular confirmation when required by the diagnostic claim, an atypical phenotype or the treatment indication. Positive screening IHC is not automatically a confirmed fusion or mutation.

Trust beside the reviewed antibody

A Trust button beside a reviewed marker opens its assessed uses, the reason for the judgment, supporting references and when confirmation is needed. One antibody may have several assessments for different organs, specimens or clinical questions. An assessment for one use does not certify every row carrying that antibody.

Defined use identifies a specified guideline or assay/indication context. Restricted adjunct requires a panel and the stated limits. Screen; confirm requires the appropriate confirmatory pathway. Not for this claim rejects a particular inference, not the antibody’s other uses. No button means that no assessment from this approved review is attached; it does not mean the marker is unreliable.

The assessments are editorial judgments. They are not numerical trust scores, formal GRADE ratings, local assay validation or approval of historical numerical estimates. Diagnostic ranking weights remain unchanged pending their own calibration and validation.

How we judge the supporting paper

The review unit is a defined marker–claim–assay–clinical setting. We read the methods and relevant supplements, reconstruct the evaluable denominator, inspect the reference diagnosis and examine important mimics. We record exclusions, failed tests, incomplete molecular verification and cohort overlap.

FrameworkQuestion it addresses
CAP, 2024Has the assay been analytically validated or verified for its intended use? Analytic concordance is distinct from clinical sensitivity or specificity.
QUADAS-3, 2026What are the risk-of-bias and applicability concerns for the diagnostic-accuracy estimate?
STARD, 2015Are methods and results sufficiently reported to assess the claim? Complete reporting does not itself establish low risk of bias.
GRADEHow certain is the body of evidence for the defined question? This requires an explicit evidence-body assessment, beyond a single paper.

These frameworks inform different parts of review. Workbench editorial decisions do not imply that complete formal QUADAS-3 or GRADE assessments have been performed for every marker.

What earns acceptance?

Editorial decisionWhat it means
Accept within scopeDirectly applicable evidence supports the defined use with suitable precision and corroboration or authoritative guidance. Publish the assay, clinical context and source.
Accept with restrictionsThe evidence supports a narrower application, specimen, differential or pitfall. Place the restriction beside the marker.
HoldDecisive methods are missing, counts conflict, thresholds are exploratory, or validation/applicability is inadequate. Do not turn the claim into an established quantitative rule.
Do not adopt the claimThe claimed use exceeds the study design, is contradicted or has become obsolete. Other uses of the same marker may still be supported.

These are Workbench editorial categories, not an official grading scale. There is no universal requirement for two papers, 100 cases, a prestigious journal or 90% clinical sensitivity. A case report can reveal a possible pitfall without estimating its frequency. Do not add domain judgments into a numerical “quality score” or convert these categories into diagnostic weights.

Before quoting sensitivity or specificity

Example: AKR1B10 in a renal differential

AKR1B10 positivity can occur in FH-deficient, papillary and TFE3-rearranged RCC. The main paper and supplement support a restricted adjunctive note; double AKR1B10/2SC positivity alone does not establish FH deficiency. Unresolved accuracy figures remain excluded from quantitative adoption.

Zheng et al., 2023 · Current Workbench wording

Example: MDM2 IHC versus amplification

Nuclear protein staining and gene amplification answer different questions. MDM2 IHC is a restricted adjunct; neither positivity nor negativity establishes amplification status. A confirmed amplification still requires a morphologic and clinical differential. Equivocal or unevaluable results are not negative results.

Both MDM2 entries are deliberately unscored. A trust assessment describes the supported clinical use; it does not validate a numerical weight. Apply the laboratory-validated assay and resolve uncertainty when the distinction matters.

Gambella et al., 2023 — FISH pitfalls · Machado et al., 2022 — negative IHC · MDM2 amplification in gastric cancer

Current MDM2 decisions and holds

Document a conclusion that can be checked

Record the result, assay/clone, specimen, localization, extent, controls and relevant limitations. In literature review, also record the exact supported claim, source, reading depth, review date and reason for acceptance or hold. Request the full text or supplement when it contains information needed for the decision.

The broad marker review remains in progress. An existing Workbench entry is not, by itself, evidence that its every use or numerical association has completed appraisal. See the current approved changes.